Hardware Organized by Material Type
Liquids (Yield Stress < 100 Pa)
Air-Displacement Pipettes
Principle: Movable piston creates air gap above liquid; changes in air volume precisely push/pull liquid.
Accuracy: ±1–2% (excellent)
Speed: 30 sec per 10 mL
Cost: \(200–\)2000 per unit
Vendors: Eppendorf, Rainin, Gilson, Hamilton
Suitable materials: Water, solvents, dilute solutions (< 50 cP)
Limitations: Fail at viscosity > 100 cP; cannot handle yield-stress materials
Syringe Pumps
Principle: Motor drives syringe plunger at controlled speed; linear displacement → precise volume flow.
Accuracy: ±2–5%
Speed: 5–30 sec per 10 mL (depends on speed setting)
Cost: \(1k–\)10k per pump
Vendors: New Era, Infuse/Withdraw, Aladdin
Suitable materials: Low- to medium-viscosity liquids, some semi-solids
Advantages: Versatile, works on bench top, good for research
Peristaltic Pumps
Principle: Rotating rollers squeeze flexible tubing; material never touches pump.
Accuracy: ±5–10% (depends on tubing wear)
Speed: Tunable, 1–1000 mL/min
Cost: \(500–\)5k
Vendors: Watson-Marlow, Cole-Parmer, Gilson
Suitable materials: Any liquid (food-grade, sterile, corrosive); semi-solids at low rates
Advantages: No contact, aseptic, continuous flow
Acoustic Droplet Ejection (ADE)
Principle: Focused acoustic transducer creates pressure pulse; ejects nano-droplets from well plate.
Accuracy: ±15% (droplet-scale, high relative error)
Speed: Milliseconds per droplet
Cost: \(50k–\)200k (equipment)
Vendors: Echo (Beckman Coulter), Labcyte
Suitable materials: Aqueous solutions, low-viscosity liquids
Advantages: Contactless; useful for fragile reagents (proteins, cells)
Limitations: Expensive; droplet size ~50 nL; not for viscous liquids
Semi-Solids (Yield Stress 100 Pa – 100 kPa)
Screw Extrusion (Stepper Motor)
Principle: Stepper motor drives rotating screw; screw flutes push material out.
Accuracy: ±2–5% (depends on material uniformity)
Speed: 10–60 sec per mL
Cost: \(3k–\)15k
Vendors: Asymptotic, Ultimaker (3D printer heads)
Suitable materials: Bio-inks, pastes, gels, highly viscous liquids
Advantages: Decouples flow from applied pressure; good control
Air-Pressure Syringes
Principle: Compressed air pushes on syringe plunger; material is extruded through nozzle.
Accuracy: ±5–15% (sensitive to pressure fluctuations)
Speed: 5–30 sec per mL
Cost: \(2k–\)8k (syringe + regulator)
Vendors: EFD (Eaton), Asymptotic, Nordson
Suitable materials: Pastes, gels, gentle on cells (low shear)
Advantages: Gentle, no motor vibration; good for bio-inks
Limitations: Pressure fluctuation → volume variation; requires fast feedback for ±1%
Progressive Cavity Pumps
Principle: Single rotating screw in elliptical cavity; creates expanding cells that push fluid.
Accuracy: ±1–3%
Speed: Tunable, 10–1000 mL/min
Cost: \(5k–\)20k
Vendors: Mono (Halliburton), Netzsch
Suitable materials: High-viscosity, yield-stress, shear-sensitive materials
Advantages: Gentle pumping action; excellent for viscous polymers, cells
Powders (Yield Stress 10 kPa – 1 MPa)
Auger / Screw Feeders
Principle: Rotating screw conveys powder from hopper to exit; volume per revolution ≈ constant.
Accuracy: ±2–8% (depends on flowability)
Speed: 10–30 sec per gram
Cost: \(2k–\)10k
Vendors: Maersk, Acrison, Schenck Process
Suitable materials: Free-flowing to moderately cohesive powders
Limitations: Fails with bridging powders; accuracy limited by slip
Improvement: Variable-pitch screws, vacuum assist
Vibratory Hoppers
Principle: Vibrating bin loosens powder; gravity feeds through metered orifice.
Accuracy: ±5–15% (poor for cohesive materials)
Speed: Fast, 2–10 sec per gram
Cost: \(1k–\)5k
Vendors: Vibra, Carr Lane
Suitable materials: Free-flowing powders only
Limitations: Electrostatic buildup; poor accuracy; doesn't work for sticky powders
Vacuum-Assisted Dosing
Principle: Suction creates vacuum through metered orifice; powder is pulled out by air flow.
Accuracy: ±1–3%
Speed: 10–20 sec per gram
Cost: \(5k–\)15k
Vendors: GEA, Schenck Process, Coptis
Suitable materials: Cohesive and fine powders; even electrostatic materials
Advantages: Works with difficult powders; good accuracy
Fluidized-Bed Dispensers
Principle: Air flows up through porous membrane; powder floats weightless; exit through calibrated nozzle.
Accuracy: ±0.5–2% (excellent)
Speed: 10–30 sec per gram
Cost: \(10k–\)50k
Vendors: Coptis, Netzsch, Schenck Process
Suitable materials: All powders (even bridging, cohesive, electrostatic)
Advantages: Highest accuracy; works with all powder types
Limitations: High cost; requires compressed air supply
Gravimetric Feedback Systems
Principle: Real-time weighing on collection vessel; adjusts feeder speed to match target.
Accuracy: ±0.5–1% (depends on scale precision)
Speed: 20–60 sec per dose (includes feedback cycles)
Cost: \(8k–\)30k (feeder + scale + controller)
Vendors: Mettler-Toledo, Schenck Process, Acrison
Suitable materials: Any powder
Control strategy: Iterative Optimization (Strategy 2) or Real-Time Correction (Strategy 3)
Solids (Yield Stress > 1 MPa)
6-DOF Collaborative Robots
Principle: Motorized joints, each with encoders and torque sensors; compute target positions via kinematics.
Accuracy: ±0.03–0.05 mm (excellent positional repeatability)
Speed: Cycle time 5–30 sec (depends on task)
Cost: \(30k–\)80k per arm
Vendors: Universal Robots (UR3e, UR5), KUKA, Franka Emika, ABB
Suitable for: Object pick-and-place, assembly, inspection
Control strategies: Tactile Retrieval (Strategy 7), Real-Time Correction (Strategy 3)
Grippers
Parallel Gripper
- Cost: \(1k–\)3k
- Grip force: 100–1000 N
- Suitable: Rigid objects, boxes, rigid parts
- Limitations: Cannot adapt to soft or irregular shapes
Vacuum Gripper
- Cost: \(500–\)2k (suction cup only)
- Lifting capacity: 1–50 kg (depends on cup area, vacuum pump)
- Suitable: Flat, smooth, non-porous objects
- Limitations: Won't work on porous or deformable materials
Soft Gripper
- Cost: \(2k–\)10k
- Grip force: 1–10 N (gentle)
- Suitable: Delicate objects (tissue, fruit, fragile electronics)
- Limitations: Lower speed; may slip on smooth hard objects
Electroadhesion Gripper
- Cost: \(3k–\)10k
- Lifting capacity: 10–500 g (depends on voltage, material)
- Suitable: Any material (uses electrostatic force)
- Advantages: Non-contact; works on glass, foil, ceramics
- Limitations: Requires conductive gripper pad and target
Pick-and-Place Systems (Gantry / SCARA)
Principle: Fast XY(Z) motion; specialized for high-speed discrete part handling.
Accuracy: ±0.5–2 mm
Speed: 1–3 sec per pick (very fast)
Cost: \(15k–\)80k
Vendors: ABB, Stäubli, Fanuc, Yaskawa
Suitable for: Tablets, pellets, small assemblies (food, pharma, electronics)
Advantages: Faster than 6-DOF robots for planar tasks
Comparison Table: Key Commercial Systems
| System | Yield Stress Range | Accuracy | Speed | Cost | Strategy |
|---|---|---|---|---|---|
| Air-displacement pipette | 1–50 Pa | ±1–2% | 30 sec/10mL | $1k | Characterize |
| Syringe pump | 10–1000 Pa | ±2–5% | 10 sec/mL | $5k | Iterate |
| Air-pressure syringe | 100–10k Pa | ±5–15% | 10 sec/mL | $5k | Real-time correct |
| Screw extrusion | 500–50k Pa | ±2–5% | 20 sec/mL | $8k | Iterate/Predict |
| Auger feeder | 10k–1M Pa | ±2–8% | 10 sec/g | $5k | Characterize |
| Vacuum dispenser | 100k–1M Pa | ±1–3% | 15 sec/g | $10k | Iterate |
| Fluidized-bed | 100k–10M Pa | ±0.5–2% | 20 sec/g | $30k | Iterate/Feedback |
| Collaborative robot | > 1M Pa | ±0.03 mm | 5–30 sec | $50k | Tactile |
Further Exploration
For each material type, see the Material Properties page for characterization protocols, and Recurring Strategies for recommended control approaches.
For cost vs. accuracy trade-offs, see By Yield Stress & Cost.